Install Test
Return Cables
When all parts are ready for use, connect the test port
extension cables. To connect a cable to the test set:
Loosen the test port anti-rotation clamp securing screw
and slide the clamp over the cable far enough to provide
access to the cable connector.
Connect the cable to the test port and tighten using the
torque wrench.
Slide the clamp toward the port, aligning it so that the
flats on the clamp mate with flats around the test port.
Tighten the clamp securing screw. This clamp keeps the
cable from becoming loose from the test set connector.
Install Precision Adapters
T
o install the precision adapters and adapter anti-rotation
clamp:
. Connect the adapter to the cable, tighten finger tight, and
then use the spanner and the torque wrench to achieve
the final torque.
l
Loosen the adapter anti-rotation clamp securing screws
and slide the clamp over the adapter. Align the clamp so
that it can grip both the cable connector body and the
adapter body. Tighten the clamp securing screws. The
screw tightening order is not important.
The adapter anti-rotation clamp assures that the TRL
adapter does not become loose during calibration and
measurement.
l
Connect the second TRL adapter and its anti-rotation
clamp to serve as Port 2.
PORT 1
P O R T 2
Two Cable Setup with Components Connected
16