Adapter Removal
Calibration Procedure
Create Cal Set for Port
1
Port
Cal
Srt
Create Cal Set for Port 2
Port
2 Cal
Adapter Removal Mathematics
Adapter and
the
First create the Port 1 calibration set by performing the TRL
calibration between Port 1 and the adapter. Keep
the adapter connected to Port 2 during the entire
procedure. Save the calibration in Cal Set
Next create the Port 2 calibration set by performing the
TRL 2-Port calibration between Port 2 and the adapter.
Keep the adapter connected to Port 1 during the entire
procedure. Save the calibration in Cal Set 2.
The Adapter Removal function mathematically combines
the Port 1 cal set with the Port 2 cal set to produce a third
cal set having the effects of the adapter removed. The
resulting cal set is as if Port 1 and Port 2 could be connected
together.
When the two calibrations have been saved, proceed with
the adapter removal sequence as follows:
Press:
CAL
MORE
MODIFY CAL SET
ADAPTER REMOVAL
Press
CAL SET for PORT 1,
then
CAL SET 1.
Press
CAL SET for PORT 2,
then
CAL SET 2.
Press
ADAPTER 3.5 mm C.l.
This specifies the
calibration kit that includes the length specification for the
precision adapter.
Press
MODIFY SAVE,
then
CAL SET 3.
The new calibration set is computed and stored. 2-Port
correction is turned on.
Now remove the third adapter and connect the device under
test.
The display should now show an accurate S-parameter
measurement of the device.
27