Calibration Kit Components
Air Lines
The HP
3.5 mm Precision Calibration Kit contains
components useful for several different calibration tech-
niques. The parts used for the TRL 2-Port calibration are:
l
Longer precision air line for 2 to 7
l
Shorter precision air line for 7 to 32
l
Male and female short circuits
l
Male and female fixed loads
Other parts included are:
l
Three precision TRL adapters
. Two adapter anti-rotation clamps
l
Calibration kit definition tape
l
Male and female open circuits
l
Tools
Note the reference card in the box and the material list on
the bottom of the box.
These are precision air lines. Each one is always stored in
its installation tool to help prevent damage and preserve
cleanliness.
The shorter air line is used for measurement calibration
over the 7 to 32
frequency range. The longer air line is
used for measurement calibration over the 2 to 7
range. Each air line is fully specified at all frequencies
within its stated range.
7 - 3 2
2 - 7 L I N E
Line Standards 7 to 32
and 2 to 7
Air
Lines Installed in Tools with Storage Bottles
Polar Display Showing Phase Response
of
Longer
Air Line Over 2
to
7
Frequency Range
Polar Display Showing Phase Response of Shorter
Air Line over 7 to 26.5
Frequency Range