l
Move the nut forward to engage the threads, tighten
finger tight, then use the spanner and the torque wrench
to achieve the final torque.
Connect Port 2
.
Press 2-7 LINE. The HP 8510 makes six measurements.
All frequencies of the current range are measured, but only
data from 1
to 7
is used to compute error
coefficients.
To remove the 2-7
air line:
l
Loosen the air line nut and carefully move Port 2 away
from Port 1.
Disconnect Port 2
Engage
the Tool
TURN
HOLD
l
Insert the tool and lightly press it to engage the center
conductor.
l
Loosen the adapter nut and then disconnect the outer
conductor from the adapter. The center conductor will be
removed with the outer conductor.
Remove 2-7
Line
Note: It is important to be very careful to move Port 2 away
in a straight motion because the center conductor may stay in
Port 2. If the center conductor remains with the female port,
use the installation tool to remove it. Carefully insert the
center conductor back into the insertion tool, male end first.
Remove the Center Conductorfrom Port 2 necessary)
23