Test 7: Variable Baseline (Option 001)
Note
This
is only to be performed if
Option 001
is installed.
Test Specifications
Range:
Accuracy:
of amplitude
source impedance only
Equipment Needed
1. Digitizing Voltmeter (DVM)
2. High Power attenuator
3.
Feedthrough
Procedure
1. On the HP 8114A set up the OUTPUT page as shown in the
following illustration:
Configuring the Output Page
2. Connect HP 8114A to the DVM as shown:
8 4A
TEST
OHM
Equipment Set-up 1 for Variable Baseline
edings for the following baseline settings (VB). Multiply the
3. Set the DVM HP 3458A to:
Function: DCV
Trigger: TRIG EXT
AD-Converter integration time NPLC: 0.1
(NPLC = Number of Power Line Cycles)
Testing the HP 8114A