HP 8114A Service Guide
-314
memory lost
HP-IB message =
Setting [no] Firmware
revision nr. incompatible”
Local message =
Setting [no] Firmware
rev. incompat.
Description:
= remote or local,
= number input]
Each firmware issue has a revision number. The
instrument is trying to read a setting with an
incompatible revision number.
memory lost
HP-IB message =
Setting [no] Error
check conflict”
Local message =
Setting [no] Error check
conflict”
Description:
= remote or local, [no] = number input]
The error check switch in the specified setting is OFF
but the internal error check is ON. The instrument
cannot recall this setting.
memory lost
HP-IB message =
Setting [no] Instrument
configuration incompatible”
Local message =
Setting [no]
incompatible”
Description:
= remote or local, [no] = number input]
Every stored setting contains the instrument
configuration. The stored configuration does not
correspond to the actual instrument configuration.
failed
HP-IB message = “Selftest error: Microprocessor
board failed”
Local message = “Selftest error:
board
failed”
Description:
Hardware on the Microprocessorboard does not work.
-314
-314
-330
Troubleshooting