A) Amplitude
(from
into
Procedure
Connect HP
the DVM as shown:
TO BNC ADAPTOR
2. On the HP 8114A set up the
page as shown in the
following illustration:
Configuring the Output Rage
3. Set HP 8114A Delay to
0
ns
4.
the baseline reading as
5. Set HP 8114A Delay to 50 ms and Amplitude to 1 V
6.
the high-level reading as
7. Calculate the Amplitude:
A
=
8. Repeat steps 3 to 7 for the following settings:
Amplitude
Amplitude limit
TR
minimum maximum
0.89 v
1.11 v
6 - l
2 v
1.38 v
2.12 v
6-2
4.86 v
5.15 v
6 - 3
10 v
9.80 v
10.2 v
6-4
20 v
19.7 v
20.3 V
6 - 5
v
49.4 v
50.6 V
6 - 6
3-10. Amplitude Levels: 503 into
7
Testing the HP