8114A Specifications
Variable Baseline Option 001
- - Maximum Duty Cycle, see Figure 6.1
Maximum Duty
15%
Figure 2-2. Baseline Duty-cycle Amplitude Ranges
Refer to Figure 2-2. Under the following conditions Figure 2-l still
applies for the maximum duty-cycle:
n
Positive pulse with negative Baseline
Negative pulse with positive Baseline
n
-5 V <Baseline + 5 negative or positive pulses
Under the following conditions maximum duty-cycle is 15%:
Baseline + 5 V and positive pulses
Baseline
V and negative pulses
Note also, that the pulse is limited to a
V window (500 into 5061)
so that for positive pulses with positive Baseline, or negative pulses
with negative baseline, the maximum available amplitude becomes
limited by the Baseline setting
HP 8114A Pulse Generator