Test 3: Delay
Test Specifications
Resolution
Accuracy
RMS-Jitter
Fixed: typical 42.0 ns
Variable: 0.00 ns to 999 ms
3 digits, best case ps
0.03% + 26
ps (0.05%
ps for 50 ns <delay
Equipment Needed
Digitzing Oscilloscope with Accessories
Pulse Generator
Counter
Cable, 50 coaxial, BNC
Procedure
1. Connect HP 8114A to the Scope as shown:
WITH
A D A P T O R
Connecting
8114A to the Scope
2. Set the Pulse Generator to:
Period
Width
100 ns
Amplitude 1 V
Offset
output
Enable
3. Select the TRIGGER page on the HP
and set up as follows:
The TRIGGER
Set-up
Testing the HP