2. On the HP 8114A set up the OUTPUT page as shown in the
following illustration:
Configuring the Output Page
3. Set the Digitizing Oscilloscope HP 54121T:
n
Press
n
Select the Display menu and set the Number of Averages to
n
Select the Channel menu and set the Attenuation factor of
channel 4 to 10
Set the
of channel 4 to 10 mV/div
Set OFFSET to 1 V
Select the
menu and set the
to 100
Center the first negative-going edge of the signal (approximate
Delay = 38.8 ns)
n
Select
the
Delta V menu and turn the V markers On
n
Set the Marker Position to 1.02 V and the Marker 2 Position to
1.00 v
n
Select the Delta t menu and turn the T Markers On
n
Set START ON EDGE =
and STOP ON EDGE =
n
Press the
(
PRECISE EDGE FIND
]
key
Testing the HP