![HP 64700 series User Manual Download Page 295](http://html.mh-extra.com/html/hp/64700-series/64700-series_user-manual_161481295.webp)
To measure longer intervals, the sample period must be increased.
Calculation
R esolution is formally defined as:
resolution = + /- (sample skew).
A high sample rate (100 MHz), coupled with low skew (plus or
minus 4 ns for both opposite and same direction transitions), gives
the timing analyzer very good resolution.
1
Improving the
Accuracy of Time
Interval
Measurements
You can improve the accuracy of an interval measurement by
making the measurement with a series of repetitive executions.
When a single execution is made, the resolution is equal to
+ /-(sample skew). When measuring a stable interval
using a series of repetitive executions, the accuracy of the
measurement improves by the following:
accuracy = + /-(((sample period)/sqrt(n)) + skew)
where:
sample period
is the sample period specified in the timing
analyzer trace specifications.
sqrt(n)
is the square root of the number of
executions included in the measurement.
skew
is the delay differences between input
channels.
A tenfold improvement (x10) is obtained in the accuracy of your
measurement when using one hundred repetitive executions for the
measurement.
Accurate Timing Measurements D-3
Summary of Contents for 64700 series
Page 2: ......
Page 19: ...Tables Table 9 1 Summary of Commands 9 4 Table 9 2 Command Assignments 9 5 Contents 11 ...
Page 20: ...1 Notes 12 Contents ...
Page 24: ...1 Notes 1 4 Introduction ...
Page 106: ...1 Notes 4 36 Performance Measurements ...
Page 118: ...Figure 5 8 True Demultiplexing 5 12 Using the External Analyzer ...
Page 128: ...1 Notes 6 6 Timing Introduction ...
Page 134: ...1 Notes 7 6 Timing Getting Started ...
Page 168: ...1 Notes 8 34 Timing Using the Analyzer ...
Page 201: ...1 find This command finds a trigger like event in trace memory Syntax Timing Commands 9 33 ...
Page 219: ...1 mark This command marks specified conditions in trace memory Syntax Timing Commands 9 51 ...
Page 254: ...1 trigger This command specifies trigger conditions Syntax 9 86 Timing Commands ...
Page 262: ...1 Notes 9 94 Timing Commands ...
Page 280: ...1 Notes B 16 Timing Diagrams and Outputs ...
Page 292: ...1 Notes C 12 Timing Messages ...