General
Specifications
Threshold Accuracy = + /- 50 mV.
Threshold Voltage R ange = -6.40V to 6.35V.
Dynamic R ange = + /- 10 V about threshold setting.
Minimum Input Swing = 600 mV pp.
Minimum Input Overdrive = 250 mV or 30% of threshold
setting, whichever is greater.
Absolute Maximum Input Voltage = + /- 40 V.
Probe Input R esistance = 100K ohms + /- 2% .
Probe Input Capacitance = approximately 8 pF.
M 5 V Probe Current = 0.650 A.
+ 5 V Probe Voltage Accuracy = + 5.0 + /- 5% .
State Analyzer
Data Setup Time = 10 ns min.
Data Hold Time = 0 ns min.
Qualifier Setup Time = 20 ns min.
Qualifier Hold Time = 5 ns min.
External Analyzer Specifications A-1
Summary of Contents for 64700 series
Page 2: ......
Page 19: ...Tables Table 9 1 Summary of Commands 9 4 Table 9 2 Command Assignments 9 5 Contents 11 ...
Page 20: ...1 Notes 12 Contents ...
Page 24: ...1 Notes 1 4 Introduction ...
Page 106: ...1 Notes 4 36 Performance Measurements ...
Page 118: ...Figure 5 8 True Demultiplexing 5 12 Using the External Analyzer ...
Page 128: ...1 Notes 6 6 Timing Introduction ...
Page 134: ...1 Notes 7 6 Timing Getting Started ...
Page 168: ...1 Notes 8 34 Timing Using the Analyzer ...
Page 201: ...1 find This command finds a trigger like event in trace memory Syntax Timing Commands 9 33 ...
Page 219: ...1 mark This command marks specified conditions in trace memory Syntax Timing Commands 9 51 ...
Page 254: ...1 trigger This command specifies trigger conditions Syntax 9 86 Timing Commands ...
Page 262: ...1 Notes 9 94 Timing Commands ...
Page 280: ...1 Notes B 16 Timing Diagrams and Outputs ...
Page 292: ...1 Notes C 12 Timing Messages ...