![HP 64700 series User Manual Download Page 25](http://html.mh-extra.com/html/hp/64700-series/64700-series_user-manual_161481025.webp)
2
Getting Started
1
Introduction
This chapter shows you how to use the emulation analyzer from
within the Softkey Interface.
This chapter describes:
The sample program on which example measurements are
made.
The default, power-up condition of the analyzer (including
how to begin the trace measurement and display the trace).
Expressions in trace command qualifiers.
This chapter shows you how to:
Specify a simple trigger (and change the trigger position).
Specify a storage qualifier.
Use trace prestore.
Change the count qualifier.
Trigger on the Nth occurrence of some state.
Trigger on multiple states.
Use the sequencer.
Stop a trace measurement.
Getting Started 2-1
Summary of Contents for 64700 series
Page 2: ......
Page 19: ...Tables Table 9 1 Summary of Commands 9 4 Table 9 2 Command Assignments 9 5 Contents 11 ...
Page 20: ...1 Notes 12 Contents ...
Page 24: ...1 Notes 1 4 Introduction ...
Page 106: ...1 Notes 4 36 Performance Measurements ...
Page 118: ...Figure 5 8 True Demultiplexing 5 12 Using the External Analyzer ...
Page 128: ...1 Notes 6 6 Timing Introduction ...
Page 134: ...1 Notes 7 6 Timing Getting Started ...
Page 168: ...1 Notes 8 34 Timing Using the Analyzer ...
Page 201: ...1 find This command finds a trigger like event in trace memory Syntax Timing Commands 9 33 ...
Page 219: ...1 mark This command marks specified conditions in trace memory Syntax Timing Commands 9 51 ...
Page 254: ...1 trigger This command specifies trigger conditions Syntax 9 86 Timing Commands ...
Page 262: ...1 Notes 9 94 Timing Commands ...
Page 280: ...1 Notes B 16 Timing Diagrams and Outputs ...
Page 292: ...1 Notes C 12 Timing Messages ...