Chapter 1 Performance Tests
HP 53181A Performance Test Record (Page 2 of 4)
1-72
Assembly-Level Service Guide
1
HP 53181A Performance Test Record
(Page 2 of 4)
Complete Performance Tests
Test
Number
Test Description
Minimum
Actual Reading
Maximum
1
Trigger Level:
Channel 1
−
15 mV
−
15 mV
1. __________
2. __________
+15 mV
+15 mV
Test 2: Channel 1 Frequency Sensitivity (HP 53181A Only)
Standard (Front Panel Terminals)
Actual Reading
Specification
100 kHz–100 MHz, Channel 1
1. __________
≤ −
21 dBm (20 mVrms)
100 MHz–200 MHz, Channel 1
2. __________
≤ −
17.5 dBm (30 mVrms)
200 MHz–225 MHz, Channel 1
3. __________
≤ −
15.0 dBm (40 mVrms)
Option 060 Rear Terminals
Actual Reading
Specification
100 kHz–225 MHz, Channel 1
1. __________
≤ −
9.5 dBm (75 mVrms)
Test
Number
Test Description
Minimum
Actual
Reading
Maximum
3
Channel 1
Frequency
Accuracy
(HP 53181A Only)
1 MHz Test
100 MHz Test
200 MHz Test
225 MHz Test
999,999.999 Hz
99.999,999,93 MHz
199.999,999,87 MHz
224.999,999,85 MHz
1. __________
2. __________
3. __________
4. __________
1, 000, 000.001 Hz
100.000, 000,07 MHz
200.000,000,13 MHz
225.000,000,15 MHz
Summary of Contents for 53131A
Page 1: ...Assembly Level Service Guide HP 53131A 132A 181A 225 MHz Counter ...
Page 2: ......
Page 14: ...Contents xii Assembly Level Service Guide ...
Page 22: ...Preface xx Assembly Level Service Guide ...
Page 23: ...1 Performance Tests Verifying Specifications ...
Page 97: ...2 Service ...
Page 149: ...3 Replacing Assemblies Disassembly and Reassembly ...
Page 173: ...4 Retrofitting Options ...
Page 191: ...5 Replaceable Parts ...
Page 201: ...6 Backdating Manual Changes ...
Page 218: ...Chapter 6 Backdating Backdating Specifications 6 18 Assembly Level Service Guide 6 ...
Page 219: ...7 HP 53131A 132A Specifications ...
Page 235: ...8 HP 53181A Specifications ...
Page 246: ...Chapter 8 HP 53181A Specifications General Information 8 12 Assembly Level Service Guide 8 ...
Page 252: ......
Page 254: ... Manual Part Number 53131 90039 Printed in U S A MARCH 1999 ...