Chapter 1 Performance Tests
Test 5: Option 030/050/124 Channel 3 Frequency Sensitivity (HP
53131A/132A Only)
Assembly-Level Service Guide
1-33
1
200 MHz to 5.0 GHz Sensitivity for HP 53131A/132A
(Option 050 only)
1
Connect the 8340B Sweep Oscillator to Channel 3 of the counter.
2
Change the settings of the Sweep Oscillator to 5.0 GHz,
−
40 dBm.
3
Increase the power level in 1 dB steps until the Counter displays a
stable reading of approximately 5.0 GHz.
The Counter should display 5.0 GHz at
≤ −
23 dBm.
NOTE
Use a power meter and sensor to ensure the accuracy of the power being
input.
4
Sweep the frequencies from 200 MHz to 5.0 GHz.
The Counter should read frequencies from 200 MHz to 5.0 GHz at
≤ −
3 dBm.
5
Record the actual reading in the Performance Test Record
(Test 5, Line 3).
6
Disconnect the test setup.
Summary of Contents for 53131A
Page 1: ...Assembly Level Service Guide HP 53131A 132A 181A 225 MHz Counter ...
Page 2: ......
Page 14: ...Contents xii Assembly Level Service Guide ...
Page 22: ...Preface xx Assembly Level Service Guide ...
Page 23: ...1 Performance Tests Verifying Specifications ...
Page 97: ...2 Service ...
Page 149: ...3 Replacing Assemblies Disassembly and Reassembly ...
Page 173: ...4 Retrofitting Options ...
Page 191: ...5 Replaceable Parts ...
Page 201: ...6 Backdating Manual Changes ...
Page 218: ...Chapter 6 Backdating Backdating Specifications 6 18 Assembly Level Service Guide 6 ...
Page 219: ...7 HP 53131A 132A Specifications ...
Page 235: ...8 HP 53181A Specifications ...
Page 246: ...Chapter 8 HP 53181A Specifications General Information 8 12 Assembly Level Service Guide 8 ...
Page 252: ......
Page 254: ... Manual Part Number 53131 90039 Printed in U S A MARCH 1999 ...