Chapter 1 Performance Tests
HP 53131A/132A/181A Operational Verification
1-12
Assembly-Level Service Guide
1
3
Set the HP 8663A to output a 200 MHz, 100 mV sine wave signal to
Channel 1.
4
On the Counter, Press Run key.
5
Send a single pulse by pressing the MAN key on the HP 8130A, and
observe that with each press the
Gate
annunciator flashes.
The Counter should display approximately 1000.
6
For HP 53131A/132A, mark Pass or Fail in the HP 53131A/132A
Performance Test Record on page 1-41, Test 7.
7
Disconnect the test setup.
External Timebase Tests
These tests verify the External Timebase 1, 5, 10 MHz specifications of
the HP 53131A and HP 53181A Counters by checking that TIMEBASE
AUTO allows the Counter to be synchronized to the external reference.
Note that the HP 53132A only operates with the 10 MHz external
timebase; therefore, perform the procedure in the section titled
“10 MHz External Timebase Input for the HP 53132A” on page 1-15
to verify that TIMEBASE AUTO allows the Counter to be synchronized to
the external reference.
Equipment
HP 3325B Synthesizers (2)
HP 10100C 50W Feedthrough
HP 10503 BNC Cables (2)
Summary of Contents for 53131A
Page 1: ...Assembly Level Service Guide HP 53131A 132A 181A 225 MHz Counter ...
Page 2: ......
Page 14: ...Contents xii Assembly Level Service Guide ...
Page 22: ...Preface xx Assembly Level Service Guide ...
Page 23: ...1 Performance Tests Verifying Specifications ...
Page 97: ...2 Service ...
Page 149: ...3 Replacing Assemblies Disassembly and Reassembly ...
Page 173: ...4 Retrofitting Options ...
Page 191: ...5 Replaceable Parts ...
Page 201: ...6 Backdating Manual Changes ...
Page 218: ...Chapter 6 Backdating Backdating Specifications 6 18 Assembly Level Service Guide 6 ...
Page 219: ...7 HP 53131A 132A Specifications ...
Page 235: ...8 HP 53181A Specifications ...
Page 246: ...Chapter 8 HP 53181A Specifications General Information 8 12 Assembly Level Service Guide 8 ...
Page 252: ......
Page 254: ... Manual Part Number 53131 90039 Printed in U S A MARCH 1999 ...