Chapter 2 Service
The HP 53131A/132A Calibration Procedures
2-18
Assembly-Level Service Guide
2
To Calibrate the Gain for Channels 1 and 2
1
Press any one of the arrow keys until
CAL: GAIN 1 ?
is displayed.
2
Press Enter key.
A scrolling message is displayed. Follow the instruction by performing the
following steps.
3
Connect a BNC tee connector to Channel 1 of the Counter.
4
Connect the Counter, dc power supply, and digital multimeter as
shown in Figure 2-2.
Figure 2-2. Gain Calibration Setup
5
Adjust the power supply until the digital multimeter displays
+ 5.000 V ± 0.001 V.
6
Disconnect the digital multimeter from the Counter to prevent
noise from being introduced into the measurement.
7
Press Enter key.
The Counter momentarily displays
CALIBRATING
, and then it should
display
GAIN 1 PASS
.
If the fail message is displayed, refer to the troubleshooting section in this
chapter.
+
-
-
DC Power Supply
HP 53131A/132A
Counter
Digital Multimeter
BNC Tee
Summary of Contents for 53131A
Page 1: ...Assembly Level Service Guide HP 53131A 132A 181A 225 MHz Counter ...
Page 2: ......
Page 14: ...Contents xii Assembly Level Service Guide ...
Page 22: ...Preface xx Assembly Level Service Guide ...
Page 23: ...1 Performance Tests Verifying Specifications ...
Page 97: ...2 Service ...
Page 149: ...3 Replacing Assemblies Disassembly and Reassembly ...
Page 173: ...4 Retrofitting Options ...
Page 191: ...5 Replaceable Parts ...
Page 201: ...6 Backdating Manual Changes ...
Page 218: ...Chapter 6 Backdating Backdating Specifications 6 18 Assembly Level Service Guide 6 ...
Page 219: ...7 HP 53131A 132A Specifications ...
Page 235: ...8 HP 53181A Specifications ...
Page 246: ...Chapter 8 HP 53181A Specifications General Information 8 12 Assembly Level Service Guide 8 ...
Page 252: ......
Page 254: ... Manual Part Number 53131 90039 Printed in U S A MARCH 1999 ...