Chapter 1 Performance Tests
Test 3: Channel 1 Frequency Accuracy (HP 53181A Only)
Assembly-Level Service Guide
1-57
1
4
Record the actual reading in the Performance Test Record
(Test 3, Line 1).
5
Change the signal generator frequency to 100 MHz.
Verify the Counter reads 100 MHz ±70 mHz.
6
Record the actual reading in the Performance Test Record
(Test 3, Line 2).
7
Change the signal generator frequency to 200 MHz.
Verify the Counter reads 200 MHz ±130 mHz.
8
Record the actual reading in the Performance Test Record
(Test 3, Line 3).
9
Change the signal generator frequency to 225 MHz.
Verify the Counter reads 225 MHz ±150 mHz.
10
Record the actual reading in the Performance Test Record
(Test 3, Line 4).
11
Disconnect the test setup.
Summary of Contents for 53131A
Page 1: ...Assembly Level Service Guide HP 53131A 132A 181A 225 MHz Counter ...
Page 2: ......
Page 14: ...Contents xii Assembly Level Service Guide ...
Page 22: ...Preface xx Assembly Level Service Guide ...
Page 23: ...1 Performance Tests Verifying Specifications ...
Page 97: ...2 Service ...
Page 149: ...3 Replacing Assemblies Disassembly and Reassembly ...
Page 173: ...4 Retrofitting Options ...
Page 191: ...5 Replaceable Parts ...
Page 201: ...6 Backdating Manual Changes ...
Page 218: ...Chapter 6 Backdating Backdating Specifications 6 18 Assembly Level Service Guide 6 ...
Page 219: ...7 HP 53131A 132A Specifications ...
Page 235: ...8 HP 53181A Specifications ...
Page 246: ...Chapter 8 HP 53181A Specifications General Information 8 12 Assembly Level Service Guide 8 ...
Page 252: ......
Page 254: ... Manual Part Number 53131 90039 Printed in U S A MARCH 1999 ...