FUNCTIONS
The 3314A outputs Sine, Square and Triangle functions and Arbitrary waveforms
(see ARB) with a related signal from the SYNC output. When functions are Off,
only DC Offset remains.
Operating Characteristics
These operating characteristics apply to all functions.
Frequency Range
001 Hz to 19.99MHz
.001 Hz to 1MHz, Opt 001
Amplitude Range
01 mVp-p to t Ovp-p
to 30Vp-p, Opt 001
Variable Symmetry
5% to 95%, Freq
<
2MHz
DC Offset
±
5VDC, independent of Ampl
±
1 5VDC, Opt 001
Invert Functions
AC component is inverted,
DC component not affected
Function OFF
AC component
=
Ovp-p,
DC component not affected
The Haversine function is a special sub-set of the Sine function which is
especially suited to driving mechanical systems. To output a Haversine, set the
3314A controls as
Mode
N CYCLE
N
1
Frequency
sets the period of each Haversine
Trigger . . . . .
.
. . . . . . . . . .
.
, . Internal
SWITR INTVL
"
,
,
, .. sets repetition rate of individual Haversines
Phase
,
"
.
.
,
-90degrees
Symmetry
,
changes ratio of rising to falling time without
affecting the period
Offset
,
"
1/2 of AMPTD
(-1/2 of AMPTD if +90°)
60