DATA ENTRY
(cant)
Offset Range
The 3314A's DC Offset can be set from -5V to
+
5V in 2 offset ranges. To assert
Range Hold from the front panel, press the OFFSET key and then the BLUE shifted
DOWN ARROW key. To assert Range Hold from the HP-IB, send "OFRHx" where
x is the Range Number
.
Note that the offset will change when Range Hold is
asserted from the HP-IB if the current entry is outside the normal limits of the
selected range. Pressing the OFFSET key and then the BLUE shifted DOWN
ARROW key or
OFRHO" via the HP-IB removes the Range Hold function
.
1
±
1.997Y
.
OOOY
.002V
.ooov
2
±
6.00V
2.00Y
.OlV
.OOV
Symmetry
Variable symmetry is not allowed in frequencv range 8 and conversely, frequen
-
cies
2MHz can only be programmed when SYM equals 50% and the SYM light
is off. "EO 1" will be displayed when these two parameters are in conflict.
53