![horiba LAQUA-PC1500 Instruction Manual Download Page 95](http://html1.mh-extra.com/html/horiba/laqua-pc1500/laqua-pc1500_instruction-manual_2162251095.webp)
Meter Specifications
77
● Meter Specifications
Specifications
LAQUA-PC1500
pH Range
-2.00 to 16.00 pH
Resolution
0.01 pH
Accuracy
± 0.01 pH
pH Buffer Groups
USA, NIST, DIN
Calibration Points
Up to 5 (USA, NIST) / Up to 6 (DIN)
ORP Range
± 2000.0 mV
Resolution
0.1 mV (< ±1000 mV), 1 mV (≥ ±1000mV)
Accuracy
±0.3 mV (< ±1000 mV), 0.3% of reading (≥ ±1000mV)
Calibration Option
Yes (Up to ±200 mV)
Conductivity Range
0.00 – 20.00 μS
20.1 – 200.0 μS
201– 2000 μS
2.01 – 20.00 mS
20.1 – 200.0 mS
Resolution
0.01 / 0.1 / 1 µS ; 0.01 / 0.1 mS
Accuracy
±0.6% full scale; ±1.5% full scale > 18.0 mS/cm
Reference Temperature
15.0 to 30.0 °C (adjustable)
Temperature Coefficient
0.00 to 10.00% per °C (adjustable)
Cell Constants
0.070 to 13.00 (adjustable)
Calibration Points
Up to 4 (Auto) / Up to 5 (Manual)
Units
S/cm, S/m (Auto ranging)
Resistivity Range
0.000 Ω•cm to 20.0 MΩ•cm
Resolution
0.5% full scale
Accuracy
±0.6% full scale, ±1.5% full scale > 1.80 MΩ•cm
Total Dissolved Solids (TDS) Range
0 – 10.00 ppm
10.1 – 100.0 ppm
101 – 1000 ppm
1.01 – 10.00 ppt
10.1 – 100 ppt
Resolution
0.01 / 0.1 / 1 ppm ; 0.01 / 0.1 ppt
Accuracy
±0.1% full scale
TDS Curves
EN27888, 442, NaCl, Linear (0.40 to 1.00)
Salinity Range
0.0 to 100.0 ppt / 0.00 to 10.00 %
Resolution
0.1 ppt / 0.01%
Accuracy
±0.2% full scale
Salinity Curves
NaCl / Seawater
Calibration Option
Yes
Temperature Range
-30.0 to 130.0 °C / -22.0 to 266.0 °F
Summary of Contents for LAQUA-PC1500
Page 1: ...Instruction Manual Benchtop pH ORP Conductivity Meter LAQUA PC1500...
Page 2: ......
Page 4: ...II MEMO...
Page 13: ...For Your Safety XI JPN EN61326 1 EN61010 1 II 2...
Page 24: ...Product Overview 6...
Page 46: ...pH Setup 28...
Page 54: ...Conductivity Setup 36...
Page 58: ...TDS Setup 40...
Page 65: ...Data Setup 47 Conductivity Mode key key key key key key...
Page 67: ...Data Setup 49 Conductivity Mode key key key key key key key...
Page 69: ...Data Setup 51 Conductivity Mode key key key key key key key...
Page 72: ...General Setup 54 Conductivity Mode key key key key key key key...
Page 74: ...General Setup 56 Conductivity Mode key key key key key key key...
Page 76: ...General Setup 58 Conductivity Mode key key key key key key key...
Page 78: ...General Setup 60 Conductivity Mode key key key key key key key...
Page 90: ...Printout Summary 72 Appendix 2 pH mV Relative mV Printer Format Measurement...
Page 91: ...Printout Summary 73 Resistivity Conductivity...
Page 92: ...Printout Summary 74 Salinity TDS...
Page 93: ...Printout Summary 75 Printer Format Data Log...
Page 94: ...Printout Summary 76 Conductivity pH Salinity Printer Format Calibration...
Page 99: ......