![horiba LAQUA-PC1500 Instruction Manual Download Page 87](http://html1.mh-extra.com/html/horiba/laqua-pc1500/laqua-pc1500_instruction-manual_2162251087.webp)
Technical Note
69
■ Appendix
This section describes the technical information, printer formats, and specifications of the
instrument.
● Appendix 1
pH calibration can be performed according to several buffer standards. The most common
standard is the US buffer standard. The default setup is US buffer standard. Alternative
standards that can be chosen are NIST and DIN.
The pH buffers are temperature dependent i.e. the pH value changes with change in
temperature. The meter is intelligent to detect the temperature and pH value associated with
buffer when calibration is performed. It remembers all the temperature vs. pH value for all the
standards.
The pH vs. temperature values for the various standards are listed below:
< USA >
Temp. (°C)
pH 1.68
pH 4.01
pH 7.00
pH 10.01
pH 12.46
0
1.67
4.01
7.12
10.32
13.42
5
1.67
4.01
7.09
10.25
13.21
10
1.67
4.00
7.06
10.18
13.00
15
1.67
4.00
7.04
10.12
12.81
20
1.68
4.00
7.02
10.06
12.63
25
1.68
4.01
7.00
10.01
12.45
30
1.69
4.01
6.99
9.97
12.29
35
1.69
4.02
6.98
9.93
12.13
40
1.70
4.03
6.97
9.89
11.98
45
1.70
4.04
6.97
9.86
11.84
50
1.71
4.06
6.97
9.83
11.70
55
1.72
4.08
6.97
9.81
11.57
Summary of Contents for LAQUA-PC1500
Page 1: ...Instruction Manual Benchtop pH ORP Conductivity Meter LAQUA PC1500...
Page 2: ......
Page 4: ...II MEMO...
Page 13: ...For Your Safety XI JPN EN61326 1 EN61010 1 II 2...
Page 24: ...Product Overview 6...
Page 46: ...pH Setup 28...
Page 54: ...Conductivity Setup 36...
Page 58: ...TDS Setup 40...
Page 65: ...Data Setup 47 Conductivity Mode key key key key key key...
Page 67: ...Data Setup 49 Conductivity Mode key key key key key key key...
Page 69: ...Data Setup 51 Conductivity Mode key key key key key key key...
Page 72: ...General Setup 54 Conductivity Mode key key key key key key key...
Page 74: ...General Setup 56 Conductivity Mode key key key key key key key...
Page 76: ...General Setup 58 Conductivity Mode key key key key key key key...
Page 78: ...General Setup 60 Conductivity Mode key key key key key key key...
Page 90: ...Printout Summary 72 Appendix 2 pH mV Relative mV Printer Format Measurement...
Page 91: ...Printout Summary 73 Resistivity Conductivity...
Page 92: ...Printout Summary 74 Salinity TDS...
Page 93: ...Printout Summary 75 Printer Format Data Log...
Page 94: ...Printout Summary 76 Conductivity pH Salinity Printer Format Calibration...
Page 99: ......