General Setup
52
●
General Setup
Using General setup function of the meter, you can:
• Select stability mode of the meter
• Set auto shut-off time
• Select temperature unit
• Reset the meter
To set the general functions, follow the procedure detailed below:
Prerequisites
• Switch on the meter.
• Keep the meter either in pH or conductivity mode
• General setup procedure is common in both pH and EC mode with different meter screen
display based on available set up sequence.
• In the calibration mode, the auto stable (
AS
) mode is activated. Default stability setup in
measurement mode is “auto stable” (
AS
). If you like, you can change it to “auto hold” (
AH
)
or “real time” (
RT
).
• Default auto shut-off time is 30 minutes. You can set the time from ---- to 30 minutes, where
---- indicates “no auto shut-off time” has been set and meter will be “on” continuously.
• Default temperature unit is
°C
and
you can
change the unit
to
°F
.
• Default reset meter setup is
NO
.
If you like to reset the meter, you can change it to
YES
.
• Stability judgment criteria remains same for both auto stability mode and auto hold mode.
• To return to the measurement mode,
press the
key.
Note
Tip
Summary of Contents for LAQUA-PC1500
Page 1: ...Instruction Manual Benchtop pH ORP Conductivity Meter LAQUA PC1500...
Page 2: ......
Page 4: ...II MEMO...
Page 13: ...For Your Safety XI JPN EN61326 1 EN61010 1 II 2...
Page 24: ...Product Overview 6...
Page 46: ...pH Setup 28...
Page 54: ...Conductivity Setup 36...
Page 58: ...TDS Setup 40...
Page 65: ...Data Setup 47 Conductivity Mode key key key key key key...
Page 67: ...Data Setup 49 Conductivity Mode key key key key key key key...
Page 69: ...Data Setup 51 Conductivity Mode key key key key key key key...
Page 72: ...General Setup 54 Conductivity Mode key key key key key key key...
Page 74: ...General Setup 56 Conductivity Mode key key key key key key key...
Page 76: ...General Setup 58 Conductivity Mode key key key key key key key...
Page 78: ...General Setup 60 Conductivity Mode key key key key key key key...
Page 90: ...Printout Summary 72 Appendix 2 pH mV Relative mV Printer Format Measurement...
Page 91: ...Printout Summary 73 Resistivity Conductivity...
Page 92: ...Printout Summary 74 Salinity TDS...
Page 93: ...Printout Summary 75 Printer Format Data Log...
Page 94: ...Printout Summary 76 Conductivity pH Salinity Printer Format Calibration...
Page 99: ......