Maintenance and Storage
62
● Maintenance and Storage of Electrodes
This section describes an overview of the procedures for maintenance and storage of pH, ORP
and conductivity electrodes. For the detailed procedures, refer to the instruction manual for
each electrode.
● How to clean the electrodes
When the tip of an electrode (responsive membrane and liquid junction) becomes dirty, the
response time may slow or an error may occur in the measurement results. To avoid such
errors, clean the electrode. For dirt that cannot be washed off by pure water (or deionized
water), use the cleaning solution indicated below depending on the type of dirt. After cleaning,
rinse the electrode with pure water (or deionized water).
However for pH and ORP electrodes, different cleaning solutions should be used to clean
different types of dirt.
For pH Electrode
For ORP Electrode
Type of dirt
Cleaning solution
General
Diluted neutral cleaning solution
Oil
Alcohol, or diluted neutral cleaning solution
Inorganic
substance
1 mol/L HCl or electrode cleaning solution (model 220)
Protein
Cleaning solution including protein-removing enzyme (model 250)
Alkaline
Dip in 1 mol/L HCl or electrode cleaning solution (model 220) for 1h to
2 h
Type of dirt
Cleaning solution
General
Dilute neutral cleaning solution (General dishwashing liquid
works reasonably well.)
Oil
Inorganic substance
Immerse dilute nitric acid (1:1 nitric acid)
Summary of Contents for LAQUA-PC1500
Page 1: ...Instruction Manual Benchtop pH ORP Conductivity Meter LAQUA PC1500...
Page 2: ......
Page 4: ...II MEMO...
Page 13: ...For Your Safety XI JPN EN61326 1 EN61010 1 II 2...
Page 24: ...Product Overview 6...
Page 46: ...pH Setup 28...
Page 54: ...Conductivity Setup 36...
Page 58: ...TDS Setup 40...
Page 65: ...Data Setup 47 Conductivity Mode key key key key key key...
Page 67: ...Data Setup 49 Conductivity Mode key key key key key key key...
Page 69: ...Data Setup 51 Conductivity Mode key key key key key key key...
Page 72: ...General Setup 54 Conductivity Mode key key key key key key key...
Page 74: ...General Setup 56 Conductivity Mode key key key key key key key...
Page 76: ...General Setup 58 Conductivity Mode key key key key key key key...
Page 78: ...General Setup 60 Conductivity Mode key key key key key key key...
Page 90: ...Printout Summary 72 Appendix 2 pH mV Relative mV Printer Format Measurement...
Page 91: ...Printout Summary 73 Resistivity Conductivity...
Page 92: ...Printout Summary 74 Salinity TDS...
Page 93: ...Printout Summary 75 Printer Format Data Log...
Page 94: ...Printout Summary 76 Conductivity pH Salinity Printer Format Calibration...
Page 99: ......