pH Calibration
11
■ Calibration
This section describes the calibration procedures using the LAQUA-PC1500 benchtop meter,
pH and conductivity electrodes.
● pH Calibration
Calibration is necessary for accurate pH measurement. To perform pH calibration, follow the
procedure detailed below:
Prerequisites
• Clean the pH electrode with DI (deionized) water and wipe it with tissue paper.
• Switch on the meter and plug in the pH electrode.
• Prepare the buffers required for calibration.
• Keep the meter in pH measurement mode.
• Dip the pH electrode at least 3 cm in the buffer.
• Perform two-point calibration using:
pH 7 and 4 for acidic sample.
pH 7 and 10 for alkaline sample.
• Perform three-point calibration using pH 7, 4 and 10 if you are unsure of the expected
sample pH value. It is recommended to calibrate with pH 7 first.
• Default buffer setup is
BUFF USA
. If you want to change to
BUFF NIST
or
BUFF DIN
,
refer to “P1.1 Buffer Setup” on page 26.
• To abort an ongoing calibration process at any point of time, press the
key.
• It is recommended to clear the previous calibration data before performing calibration. For
erasing the calibration data, refer to “P1.2 Erase Calibration Data” on page 27.
Note
Tip
Summary of Contents for LAQUA-PC1500
Page 1: ...Instruction Manual Benchtop pH ORP Conductivity Meter LAQUA PC1500...
Page 2: ......
Page 4: ...II MEMO...
Page 13: ...For Your Safety XI JPN EN61326 1 EN61010 1 II 2...
Page 24: ...Product Overview 6...
Page 46: ...pH Setup 28...
Page 54: ...Conductivity Setup 36...
Page 58: ...TDS Setup 40...
Page 65: ...Data Setup 47 Conductivity Mode key key key key key key...
Page 67: ...Data Setup 49 Conductivity Mode key key key key key key key...
Page 69: ...Data Setup 51 Conductivity Mode key key key key key key key...
Page 72: ...General Setup 54 Conductivity Mode key key key key key key key...
Page 74: ...General Setup 56 Conductivity Mode key key key key key key key...
Page 76: ...General Setup 58 Conductivity Mode key key key key key key key...
Page 78: ...General Setup 60 Conductivity Mode key key key key key key key...
Page 90: ...Printout Summary 72 Appendix 2 pH mV Relative mV Printer Format Measurement...
Page 91: ...Printout Summary 73 Resistivity Conductivity...
Page 92: ...Printout Summary 74 Salinity TDS...
Page 93: ...Printout Summary 75 Printer Format Data Log...
Page 94: ...Printout Summary 76 Conductivity pH Salinity Printer Format Calibration...
Page 99: ......