Conductivity Setup
29
● P1 COND Setup
Using P1 COND setup function of the meter, you can:
• Set cell constant
• Select conductivity unit
• Set calibration mode
• Set temperature coefficient
• Set reference temperature
• Erase calibration data
To set the conductivity functions, follow the procedure detailed below:
Prerequisites
• Switch on the meter.
• Press the
key to keep the meter in
COND
mode.
• Default cell constant value is
1.00
and one can set a value in between
0.07 to 13.00
.
• Default conductivity unit is set as
S/cm
. One can change the unit to
S/m
.
• Default auto calibration setup is set as
ON
, but to perform manual calibration, one need to
change the setup to
OFF
.
• Default temperature coefficient is
2.00%.
One can set a value in between
0.00% to
10.00%
.
• Default reference temperature is
25.0°C
.
One can set the value in between
15.0°C to
30.0°.
• Erasing previous calibration data is recommended for accurate calibration. Default setup is
NO
but to erase the calibration data, one need to change the setup to
YES
.
To return to the measurement mode, press the
key.
Note
Tip
Summary of Contents for LAQUA-PC1500
Page 1: ...Instruction Manual Benchtop pH ORP Conductivity Meter LAQUA PC1500...
Page 2: ......
Page 4: ...II MEMO...
Page 13: ...For Your Safety XI JPN EN61326 1 EN61010 1 II 2...
Page 24: ...Product Overview 6...
Page 46: ...pH Setup 28...
Page 54: ...Conductivity Setup 36...
Page 58: ...TDS Setup 40...
Page 65: ...Data Setup 47 Conductivity Mode key key key key key key...
Page 67: ...Data Setup 49 Conductivity Mode key key key key key key key...
Page 69: ...Data Setup 51 Conductivity Mode key key key key key key key...
Page 72: ...General Setup 54 Conductivity Mode key key key key key key key...
Page 74: ...General Setup 56 Conductivity Mode key key key key key key key...
Page 76: ...General Setup 58 Conductivity Mode key key key key key key key...
Page 78: ...General Setup 60 Conductivity Mode key key key key key key key...
Page 90: ...Printout Summary 72 Appendix 2 pH mV Relative mV Printer Format Measurement...
Page 91: ...Printout Summary 73 Resistivity Conductivity...
Page 92: ...Printout Summary 74 Salinity TDS...
Page 93: ...Printout Summary 75 Printer Format Data Log...
Page 94: ...Printout Summary 76 Conductivity pH Salinity Printer Format Calibration...
Page 99: ......