![horiba LAQUA-PC1500 Instruction Manual Download Page 63](http://html1.mh-extra.com/html/horiba/laqua-pc1500/laqua-pc1500_instruction-manual_2162251063.webp)
Data Setup
45
● Data Setup
Using Data setup function of the meter, you can:
• Set data log interval
• Print data log
• Erase data log
To set the data functions, follow the procedure detailed below:
Prerequisites
• Switch on the meter.
• Keep the meter either in pH or conductivity mode.
• Data set up procedure is common in both pH and EC mode with different meter screen
display based on available set up sequence.
• Default data log interval is “----” which indicates no data log interval has been set. Data log
interval can be set from 2 to 999 seconds.
To return to the measurement mode, press the
key.
Note
Tip
Summary of Contents for LAQUA-PC1500
Page 1: ...Instruction Manual Benchtop pH ORP Conductivity Meter LAQUA PC1500...
Page 2: ......
Page 4: ...II MEMO...
Page 13: ...For Your Safety XI JPN EN61326 1 EN61010 1 II 2...
Page 24: ...Product Overview 6...
Page 46: ...pH Setup 28...
Page 54: ...Conductivity Setup 36...
Page 58: ...TDS Setup 40...
Page 65: ...Data Setup 47 Conductivity Mode key key key key key key...
Page 67: ...Data Setup 49 Conductivity Mode key key key key key key key...
Page 69: ...Data Setup 51 Conductivity Mode key key key key key key key...
Page 72: ...General Setup 54 Conductivity Mode key key key key key key key...
Page 74: ...General Setup 56 Conductivity Mode key key key key key key key...
Page 76: ...General Setup 58 Conductivity Mode key key key key key key key...
Page 78: ...General Setup 60 Conductivity Mode key key key key key key key...
Page 90: ...Printout Summary 72 Appendix 2 pH mV Relative mV Printer Format Measurement...
Page 91: ...Printout Summary 73 Resistivity Conductivity...
Page 92: ...Printout Summary 74 Salinity TDS...
Page 93: ...Printout Summary 75 Printer Format Data Log...
Page 94: ...Printout Summary 76 Conductivity pH Salinity Printer Format Calibration...
Page 99: ......