6 Maintenance and Troubleshooting
6.2 Troubleshooting
ES-51
87
●
The measured value is blinking
The measured conductivity value exceeds the
measurement parameters (when conductivity value is
displayed).
Display range: 0.00 – 19.99 (when cell constant is 100
m
-1
)
Check this point
As shown in the diagram, use a jumper wire or bent
paper clip to short the meter by touching both the center
pin and some metal part in the electrode connector.
If the flaching measured value disappears when this
done, the meter is nomal.
Cause
How to solve problem
The sample
solution is
inappropriate.
Change to a sample solution with
properties within the measurement
range.
The electrode
cable has been
severed.
Replace the electrode.
The main body
of the meter is
defective.
Check the point described below.