6 Maintenance and Troubleshooting
6.2 Troubleshooting
86
HORIBA
●
The indicated value does not change, or there is
absolutely no response
Cause
How to solve problem
The key-lock
function is ON.
Turn the power OFF, and then turn it
back ON again.
The system is
locked.
Turn the power OFF, and then turn it
back ON again.
The electrode
connector is not
attached
correctly.
Attach the electrode connector
correctly.
Conductivity
meter is
defective.
Contact your local HORIBA
distributor.