Temperature zero adjustment [item No. 02]
This mode uses a known temperature to calibrate the
temperature compensation value. This mode is used
when calibrating the temperature of the thermometer.
1.
Immerse the electrode in a liquid with a known
temperature, and set the temperature using the
▲
and
▼
keys.
Setting range: 0.0 to 100.0 °C
The temperature sensor attached to the electrode
maintains an accuracy of
±
1
°
C, even without calibration.
The above mode should be used when a greater
precision than
±
1
°
C is required.
2.
Use the MODE key to proceed to Automatic
power-off setting (item No. 03).
HINT!
Entering the Maintenance mode
Item No. 02
Item No.
selection
Maintenance mode
Measure-
ment mode