6 Maintenance and Troubleshooting
6.2 Troubleshooting
84
HORIBA
6.2.2 More troubleshooting
This section explains how to respond to various
symptoms of trouble that are not indicated by an error
number.
●
Nothing shows up on the display when the power is
turned ON
●
The indicated value fluctuates
When there is a problem with the electrode...
When there is a problem with the main unit of the
meter...
Cause
How to solve problem
No batteries
Place batteries in the meter.
The batteries
are loaded with
the poles
reversed.
Re-insert the batteries with the
poles correctly oriented.
The battery
voltage is low.
Remove the old batteries and
correctly insert new dry-cell
batteries.
Or connect the unit to the optional
AC adapter.
Cause
How to solve problem
There are air
bubbles on the
electrode.
Shake the electrode to remove the
air bubbles.
Cause
How to solve problem
There is a motor
or other device
causing
electrical
interference.
Move the meter to a place where it
is not subject to dielectric effects.
Be sure to ground devices that are
using commercial electricity.