12
R
insing the electrode
Make sure that no dirt and other foreign matters have
adhered to the electrode before measurement.
If dirt and other foreign matters have adhered to the
electrode, rinse the electrode as described in the following
procedure.
1
Rinse the electrode with pure water.
Make sure that the membrane part has been
well rinsed with pure water.
2
Wipe off the moisture.
Wipe off the moisture using soft and clean lint-free paper or cloth.
3
Make sure that there are no dirt
adhered to the area around the
membrane.
If dirt is adhered, the measurement may not be performed
accurately. Rinse the electrode with pure water again.
View video
View video
Dirt
Note
Note