33
Electrode Preparation
Calibration
Measurement
After Measurement
3
Wipe off the moisture.
Wipe off the moisture using soft and clean lint-free paper or cloth.
4
Soak the electrode in the sample
solution prepared in step 1.
Soak the membrane cap fully into the sample
solution.
5
Shake the electrode to remove any air bubbles remaining
on the membrane.
Do not touch the stir bar with the electrode. It may damage
the electrode.
It is recommended to place the electrode in the electrode stand.
Note
Note
Tip
Tip
Shake the
electrode