![horiba 5002S-10C Instruction Manual Download Page 33](http://html1.mh-extra.com/html/horiba/5002s-10c/5002s-10c_instruction-manual_2162179033.webp)
21
Electrode Preparation
Calibration
Measurement
After Measurement
7
Soak the electrode in the first
solution for calibration prepared in
step 3.
Soak the membrane cap fully into the solution
for calibration.
Conditioning of the electrode
Soaking the electrode in a internal solution
is called conditioning. In the following
cases, perform the conditioning for 15
minutes or longer before calibration.
Before the first use
Before use after storage for one week or
longer
After the replacement of the membrane cap
If the electrode is less responsive, perform conditioning for 2
hours or longer (conditioning for 12 hours is recommended).
Do not touch the stir bar with the electrode. It may damage
the electrode.
It is recommended to place the electrode in the electrode stand.
Note
Note
Internal
solution
Note
Note
Tip
Tip
Summary of Contents for 5002S-10C
Page 1: ...Instruction Manual Ammonia Electrode 5002S 10C ELECTRODE...
Page 8: ......
Page 26: ...14...
Page 38: ...26...
Page 58: ...46...
Page 74: ...62...
Page 77: ...65 Specifications External dimensions Unit mm 16 5 O D 18 85 150...