Page 78
ST 700 Series HART/DE Option User’s Manual
Revision 4.0
9.2.3 SV Tracking
Table 18
– Maximum SV Tracking
Max SV Limit
Parameter
Description
Meterbody Temperature (SV) upper operating
limit from specification.
Set-up
None.
Max SV
Value
Parameter
Description
Highest Meterbody Temperature ever
experienced by the device.
Set-up
None - value initialized to Min SV Limit value
prior to leaving the factory. Updates to current
Meterbody Temperature automatically when
powered at user site after one minute.
NVM
Update after every 8 hours.
Time Above
Upper Stress
Limit
Parameter
Description
Accumulation of minutes that device’s
Meterbody Temperature has been above the
value of
SV Upper Stress Limit
.
Set-up
None.
NVM
Backup once each 8 hour period
Time Since
Last SV Up
Method
Description
Time that has passed since the last time
device’s Meterbody Temperature has passed
above the value of “SV Upper Stress Limit” (in
days, hours and minutes).
Set-up
None.
NVM
Backup once each 8 hour period
Table 19
– Minimum SV Tracking
Min SV Limit
Parameter
Description
Meterbody Temperature (SV) lower operating
limit from specification.
Set-up
None.
Min SV
Value
Parameter
Description
Lowest Meterbody Temperature ever
experienced by the device.
Set-up
None - value initialized to Max SV Limit value
prior to leaving the factory. Updates to current
Meterbody Temperature automatically when
powered at user site after one minute.
NVM
Update after every 8 hours.
Time Below
Lower Stress
Limit
Parameter
Description
Accumulation of minutes that device’s
Meterbody Temperature has been below the
value of “SV Lower Stress Limit”.
Set-up
None.
NVM
Backup once each 8 hour period
Time Since
Last SV
Down
Method
Description
Time that has passed since the last time
device’s Meterbody Temperature has passed
below the value of “SV Lower Stress Limit” (in
days, hours, and minutes).
Set-up
None.
NVM
Backup once each 8 hour period