Satellite XT 4-20 mA Versions
Alarm Delay
Used to define an alarm delay for concentration alarms.
Minimum value is 0, maximum setting is 99.9 seconds.
Default setting is 3.0 seconds.
Actuate the <set> key to enter and show the actual
setting.
<Actual Alarm Delay>
To change the value, use the cursor keys <left> or
<right> to reach the desired position.
Use the cursor keys <up> and <down> to select the
figures you wish to enter.
After the new value has been entered, actuate the <set>
key to confirm or <esc> to exit.
Fault Trigger
Allows the user to define, which fault conditions will acti
-
vate the Fault Relay. Ex factory the setting is Fault Only.
Use the <set> key to enter.
The instrument will show the actual setting.
Fault Only
Used to configure the Fault Relay to be activated in case
of a fault condition only.
Actuate the cursor keys <up> or <down> to change.
Actuate the <set> key to confirm or <esc> to exit.
Fault + Warning
Used to configure the Fault Relay to be activated in case
of a fault or warning condition only.
Actuate the cursor keys <up> or <down> to change.
Actuate the <set> key to confirm or <esc> to exit.
Fault + Maintenance
Used to configure the Fault Relay to be activated in case
of a fault or maintenance condition only.
Actuate the cursor keys <up> or <down> to change.
Actuate the <set> key to confirm or <esc> to exit.
All
Used to configure the Fault Relay to be activated in case
of a fault, warning, or maintenance condition.
Actuate the cursor keys <up> or <down> to change.
Actuate the <set> key to confirm or <esc> to exit.
F A U L T T R I G G E R
F AU L T ON L Y
s e l e c t :
3
o f
3
R E L A Y S
F AU L T T R I GGER
s e l e c t :
2
o f
4
F A U L T T R I G G E R
F AU L T +WARN I NG
s e l e c t :
3
o f
4
F A U L T T R I G G E R
F AU L T +MA I N T
s e l e c t :
4
o f
4
F A U L T T R I G G E R
A L L
s e l e c t :
1
o f
4
s e l e c t :
2
o f
3
R E L A Y S
A L ARM DE L A Y
p l e a s e e n t e r !
A L A R M D E L A Y
3 . 0 s e c .
6-16