Satellite XT 4-20 mA Versions
Manual K-Factor
This submenu is used to manually calibrate the
instrument by entering a new K-Factor. This factor is a
multiplication or correction value used to calibrate the
response of the instrument to a specific
gas concentration.
The factory default K-Factor for all gases is 1.00.
Use the <set> key to enter. The instrument will display
the actual KFactor.
<Actual K-Factor>
Use the cursor keys <left> or <right> to reach the
desired position. Use the cursor keys <up> and
<down> to select the figures you wish to enter. The K-
Factor must be in the range of 0.20 and 5.00.
After a new value has been entered, actuate the <set>
key to confirm or <esc> to exit.
!Note:
Individual settings for the Manual K-Factor will automatically revert to the default value 1.00,
whenever the sensor is replaced by a new sensor.
Only instruments using catalytic sensors:
Catalytic sensors are calibrated for methane. When monitoring for other combustible gases, a
K-factor must be entered.