Satellite XT 4-20 mA Versions
Alarm 1 State
Used to enable / disable the Alarm 1 activation. Default
setting is enabled. Use the <set> key to enter. The
instrument will show the actual setting.
Disabled
Allows the user to turn off the Alarm 1 activation.
Actuate the cursor keys <up> or <down> to change.
Actuate the <set> key to confirm or <esc> to exit.
Enabled
Allows the user to turn on the Alarm 1 activation.
Actuate the cursor keys <up> or <down> to change.
Actuate the <set> key to confirm or <esc> to exit.
Alarm 1 Level
Used to set individual alarm levels for Alarm 1 activa-
tion. Default setting is 1 x TLV respectively 20 % LEL for
the target gas.
Actuate the <set> key to enter and show the actual
setting.
<Actual Alarm 1 Level>
To change the value, use the cursor keys <left> or
<right> to reach the desired position. Use the cursor
keys <up> and <down> to select the figures you wish to
enter. After the new value has been entered, actuate the
<set> key to confirm or <esc> to exit.
s e l e c t :
1
o f
4
A L A R M 1
A L ARM 1 S T A T E
s e l e c t :
1
o f
2
A L A R M 1 S T A T E
D I S A B L ED
s e l e c t :
2
o f
2
A L A R M 1 S T A T E
ENA B L ED
s e l e c t :
2
o f
4
A L A R M 1
A L ARM 1 L E V E L
p l e a s e e n t e r !
A L A R M 1 L E V E L
0 . 0 5 p pm
6-5
Configuration - Section 6