MAN-08723-001 Rev. 001 page 6 of 28
Table 3: Adjudicated Review Versus Imager Review LSIL+ Descriptive Diagnosis
Summary for Each Site and All Sites Combined.
Sensitivity is a percent of “true” LSIL+ (combined LSIL, HSIL, SQ CA and GL CA) slides classified in either study
arm as LSIL+ and specificity is a percent of “true” Non-LSIL+ (combined Negative, ASCUS, AGUS) slides classified
in either study arm as Non-LSIL+.
Sensitivity
Specificity
Site/
Number
Cases
Manual
Imager
Difference
Site/
Number
Cases
Manual
Imager
Difference
Site 1
104
84.6%
(76.2, 90.9)
82.7%
(74.0, 89.4)
-1.9%
(-9.5, 5.6)
Site 1
2208
98.7%
(98.1, 99.1)
99.3%
(98.9, 99.6)
+0.6%
(0.1, 1.0)
Site 2
70.4% 72.4% +2.0%
Site 2
99.3% 98.9% -0.4%
98
(60.3, 79.2)
(62.5, 81.0)
(-6.9, 11.0)
2342
(98.8, 99.6)
(98.4, 99.3)
(-0.8, .001)
Site 3
77.4% 85.5% +8.1%
Site 3
99.2% 99.5% +0.3%
62
(65.0, 87.1)
(74.2, 93.1)
(-4.0, 20.1)
2237
(98.7, 99.5)
(99.1, 99.8)
(-0.1, 0.6)
Site 4
84.7% 78.4% -6.3%
Site 4
98.7% 98.7% -0.08%
111
(98.1, 99.1)
(76.6, 90.8)
(-14.7, 2.1)
2381
(98.2, .99.2)
(98.1, 99.1)
(-0.6, 0.4)
All
79.7% 79.2% -0.5%
All
99.0% 99.1% +0.09%
375
(75.3, 83.7)
(74.7, 83.2)
(-5.0, 4.0)
9168
(98.8, 99.2)
(98.9, 99.3)
(-0.1, 0.3)
Numbers in parentheses represent 95% confidence intervals.
The results presented in
Table 3
show that the difference between sensitivities of the
Imager Review
and
Manual Review
arms for LSIL+ for all sites combined was not statistically significant with a
95% confidence interval of –5.0% to +4.0%. The observed difference between sensitivities for
LSIL+ varied among the sites from –6.3% with a 95% confidence interval of (–14.7%; 2.1%) to
+8.1% with a 95% confidence interval of (–4.0%; 20.1%). The difference in specificity results
between the
Imager Review
and the
Manual Review
was not statistically significant with a 95%
confidence interval of -0.1% to +0.3%. The observed differences between specificities varied among
the sites from –0.4% to +0.6%.
Table 4: Adjudicated Review Versus Imager Review HSIL+ Descriptive Diagnosis
Summary for Each Site and All Sites Combined.
Sensitivity is a percent of “true” HSIL+ (combined HSIL, SQ CA and GL CA) slides classified in either study arm as
HSIL+ and specificity is a percent of “true” Non-HSIL+ (combined Negative, ASCUS, AGUS, LSIL) slides classified
in either study arm as Non-HSIL+.
Sensitivity
Specificity
Site/
Number
Cases
Manual
Imager
Difference
Site/
Number
Cases
Manual
Imager
Difference
Site 1
89.5% 92.1% 2.6%
Site 1
98.8% 99.5% +0.7%
38
(75.2, 97.1)
(78.6, 98.3)
(-8.9, 14.1)
2274
(98.3, 99.2)
(99.1, 99.8)
(0.2, 1.2)
Site 2
72.5% 70.0% -2.5%
Site 2
99.8% 99.6% -0.1%
40
(56.1, 85.4)
(53.4, 83.4)
(-15.4, 10.4)
2400
(99.5, 99.9)
(99.2, 99.8)
(-0.3, .09)
Site 3
72.7% 86.4% +13.6%
Site 3
99.7% 99.7% 0%
22
(49.8, 89.3)
(65.1, 97.1)
(-0.7, 28.0)
2277
(99.4, 99.9)
(99.4, 99.9)
(-0.2, 0.2)
Site 4
61.5% 74.4% +12.8%
Site 4
99.5% 99.8% +0.3%
39
(44.6, 76.6)
(57.9, 87.0)
(-1.7, 27.4)
2453
(99.2, 99.8)
(99.5, 99.9)
(-0.003, 0.6)
All
74.1% 79.9% +5.8%
All
99.4 %
99.6%
+0.2%
139
(66.0, 81.2)
(72.2, 86.2)
(-1.1, 12.6)
9404
(99.2, 99.6)
(99.5, 99.7)
(0.06, 0.4)
Numbers in parentheses represent 95% confidence intervals.
Summary of Contents for ThinPrep
Page 1: ...Operator s Manual ThinPrep Imager Duo Imaging Station...
Page 4: ...The ThinPrep Imaging System The ThinPrep Imaging System...
Page 5: ...Operation Summary and Clinical Information The ThinPrep Imaging System...
Page 33: ...Table of Contents Table of Contents...
Page 37: ...iv Imager Duo Operator s Manual This page intentionally left blank...
Page 38: ...1 Introduction 1 Introduction...
Page 60: ...1 INTRODUCTION 1 22 Imager Duo Operator s Manual This page intentionally left blank...
Page 61: ...2 Installation 2 Installation...
Page 66: ...3 Operation 3 Operation...
Page 99: ...4 Software Menu 4 Software Menu...
Page 172: ...5 Maintenance 5 Maintenance...
Page 180: ...5 IMAGER DUO MAINTENANCE 5 8 Imager Duo Operator s Manual This page intentionally left blank...
Page 181: ...6 Troubleshooting 6 Troubleshooting...
Page 201: ...6 TROUBLESHOOTING 6 20 Imager Duo Operator s Manual This page intentionally left blank...
Page 202: ...7 Defi nitions and Abbreviations 7 Defi nitions and Abbreviations...
Page 207: ...8 Service Information 8 Service Information...
Page 209: ...8 SERVICE INFORMATION 8 2 Imager Duo Operator s Manual This page intentionally left blank...
Page 210: ...9 Ordering Information 9 Ordering Information...
Page 213: ...10 Remote Site Server User s Guide 10 Remote Site Server User s Guide...
Page 222: ...Index Index...
Page 229: ...Operator s Manual ThinPrep Imager Duo Imaging Station...