MAN-08723-001 Rev. 001 page 11 of 28
Table 12: Unadjudicated Marginal Frequencies Summary of Specimen Adequacy
Results – All Sites Combined.
Manual Review
Imager Review
Number of Patients:
9627 9627
Descriptive Diagnosis
N % N %
Satisfactory for Evaluation
7375 76.6 7346 76.3
Satisfactory but Limited by
2186 22.7 2252 23.4
Endocervical Component Absent
1196
12.4
1397
14.5
Scant Squamous Epithelial Component
92
1.0
102
1.1
Obscuring
Blood
45 0.5 17 0.2
Obscuring Inflammation
69
0.7
68
0.7
No Clinical History
982
10.2
933
9.7
Cytolysis
4 0.0 2 0.0
Other 6
0.1
33
0.3
Unsatisfactory for Evaluation
66 0.7 29 0.3
Endocervical Component Absent
6
0.1
0
0.0
Scant Squamous Epithelial
Component
35 0.4 22 0.2
Obscuring Blood
17
0.2
2
0.0
Obscuring
Inflammation
8 0.1 5 0.1
No Clinical History
2
0.0
2
0.0
Cytolysis
0 0.0 0 0.0
Other
2 0.0 0 0.0
Note: Some patients had more than one diagnostic subcategory
.
For SAT cases, there was agreement between the
Manual Review
cases (7375) and the
Imager
Review
cases (7346). For SBLB cases, there is agreement between the
Manual Review
cases (2186)
and the
Imager Review
cases (2252). Unsatisfactory cases were greater in the
Manual Review
cases
(66) versus the
Imager Review
cases (29).
The adjudicated results were used as a “gold standard” to define “true” specimen adequacy
classifications of the Bethesda System: SAT/SBLB and UNSAT. There were 58 “true” UNSAT
cases and 9569 “true” SAT/SBLB cases.
Table 13
below summarizes specimen adequacy performance for the Imager Review and Manual
Review arms for all four sites and all sites combined using the Bethesda System 1991 criteria.
Table 13: Adjudicated Review Versus Imager Review Specimen Adequacy
Summary for All Sites and All Sites Combined.
Sensitivity is a percent of “true” UNSAT slides classified in either study arm as UNSAT and specificity is a percent of
“true” SAT/SBLB slides classified in either study arm as SAT/SBLB.
Sensitivity
Specificity
Site/
Number
Cases
Manual
Imager
Difference
Site/
Number
Cases
Manual
Imager
Difference
Site 1
0% 0% 0.0%
Site 1
100% 100% 0.0%
21
(0/21) (0/21) (0/21)
2292
(2292/2292) (2292/2292) (0/2292)
Site 2
100% 16.7% -83.3%
Site 2
98.9% 99.6% +0.6%
6
(6/6) (1/6) (-5/6)
2476
(2449/2476) (2465/2476) (16/2476)
Site 3
80.0% 60.0% -20.0%
Site 3
99.2% 99.7% +0.5%
5
(4/5) (3/5) (-1/5)
2323
(2304/2323) (2315/2323) (11/2323)
Site 4
30.8% 19.2% -11.5%
Site 4
99.9% 99.9% +0.04%
26
(8/26) (5/26) (-3/26)
2478
(2475/2478) (2476/2478) (1/2478)
All
58
CI*
29.3%
(17/58)
(18.1, 42.7)
13.8%
(8/58)
(6.1, 25.4)
-15.5%
(-9/58)
(-25.9, -5.0)
All
9569
CI*
99.5%
(9520/9569)
(99.3, 99.6)
99.8%
(9548/9569)
(99.7, 99.9)
+0.3%
(28/9569)
(0.2, 0.4)
*95% Confidence Interval
Summary of Contents for ThinPrep
Page 1: ...Operator s Manual ThinPrep Imager Duo Imaging Station...
Page 4: ...The ThinPrep Imaging System The ThinPrep Imaging System...
Page 5: ...Operation Summary and Clinical Information The ThinPrep Imaging System...
Page 33: ...Table of Contents Table of Contents...
Page 37: ...iv Imager Duo Operator s Manual This page intentionally left blank...
Page 38: ...1 Introduction 1 Introduction...
Page 60: ...1 INTRODUCTION 1 22 Imager Duo Operator s Manual This page intentionally left blank...
Page 61: ...2 Installation 2 Installation...
Page 66: ...3 Operation 3 Operation...
Page 99: ...4 Software Menu 4 Software Menu...
Page 172: ...5 Maintenance 5 Maintenance...
Page 180: ...5 IMAGER DUO MAINTENANCE 5 8 Imager Duo Operator s Manual This page intentionally left blank...
Page 181: ...6 Troubleshooting 6 Troubleshooting...
Page 201: ...6 TROUBLESHOOTING 6 20 Imager Duo Operator s Manual This page intentionally left blank...
Page 202: ...7 Defi nitions and Abbreviations 7 Defi nitions and Abbreviations...
Page 207: ...8 Service Information 8 Service Information...
Page 209: ...8 SERVICE INFORMATION 8 2 Imager Duo Operator s Manual This page intentionally left blank...
Page 210: ...9 Ordering Information 9 Ordering Information...
Page 213: ...10 Remote Site Server User s Guide 10 Remote Site Server User s Guide...
Page 222: ...Index Index...
Page 229: ...Operator s Manual ThinPrep Imager Duo Imaging Station...