-16-
Q1E-EP1234
Manual Cleaning:
Prepare the detergent solution in a tank with cold water (please
follow the instructions of the detergent manufacturer regarding
application, dilution and contact time).
A)
EUP-O732T probe
1)
The temperature of the detergent solution should be between
15-30 °C, concentration is 1.6%. Please note the minimum
contact time of the detergent in the manufacturer’s
instruction. If a differing detergent is used, please also note
the approved material compatibility for the medical device.
2)
Immerse the immersible part of the probe without connector into
the diluted detergent solution (see Fig. 13). Wipe the
immersible part of the probe under the surface of the detergent
solution with a soft cloth to remove all visible soil. Be sure
that all grooves of the probe are implemented during the
cleaning process.
3)
The immersible part of the probe should be left in the detergent
solution according to the specified contact time of the
detergent manufacturer.
4)
Wipe the un-immergible parts of the probe with a soft cloth
dipped with the detergent solution.
5)
Rinse the probe with running tap water for 1 minute.
(alternatively: immerse the immersible part of the probe in
a tray filled with deionized water/tap water (see Fig.) for
5 min.)
6)
Visually check the outer surface of the probe for cleanness.
If necessary, use magnifying glass for visually check. If there
is still soil visible, repeat all above steps.
B)
Attachment
1)
The temperature of the detergent solution should be between
15-30 °C, concentration is 1.6%. Please note the minimum
contact time of the detergent in the manufacturer’s
instruction. If a differing detergent is used, please also note
the approved material compatibility for the medical device.
2)
Immerse the Attachment into the diluted detergent solution.
Wipe it under the surface of the detergent solution with a soft
cloth to remove all visible soil. Be sure that all grooves of
the attachment are implemented during the cleaning process.
3)
The Attachment should be left in the detergent solution
according to the specified contact time of the detergent
manufacturer.
4)
Rinse the Attachment with running tap water for 1 minute.
(alternatively: immerse it in a tray filled with deionized
water/tap water (see Fig.) for 5 min.)
Summary of Contents for EUP-O732T
Page 33: ... 27 Q1E EP1234 29 26 24 Unit mm 2900 Fig 17 Dimensions ...
Page 35: ......