-24-
Q1E-EP1234
8.
Specifications
8.1
Probe
Type:
EUP-O732T Intraoperative probe
Acoustic working frequency: 7.5MHz±30%
Technology:
Convex Array Probe
Dimensions: See
Fig.17
Weight: Approx.
0.65kg (Including cable and
connector)
Probe materials:
Biocompatible allergy free
components
Acoustic output:
According to IEC 60601-2-37 (See
Main Unit manual)
Applicable systems:
Depending on production and upgrade
status. For detailed information,
contact a service support.
Classification:
MDD classification IIa.
Cleaning: Applicable
detergents are listed in
the supplier’s list
Disinfection:
Applicable disinfectants are
listed in the supplier’s list
Sterilization:
ETO gas sterilization
Plasma sterilization
Operating conditions:
Ambient temperature:
+10 - +40°C
Contact surface temperature
(Temperature of examinee):
Max. 42°C
Relative humidity:
30 - 85% (subject to no
condensation)
Atmospheric pressure:
700hPa to 1060hPa
Storage conditions:
Temperature:
-10 - +50°C
Relative humidity:
10 - 90% (subject to no
condensation)
Atmospheric pressure:
700hPa to 1060hPa
Summary of Contents for EUP-O732T