WARNING
The sterilization method and operating conditions are as follows.
Sterilization Method
Condition
Plasma Sterilization:
STERRAD® 50, 100S or 200 (*) Short Cycle
Plasma Sterilization:
Sterrad® NX or 100NX (*)
Standard cycle
ETO Sterilization
Gas Type: 10% EO/ 90% HCFC
Temperature: 50-55°C
Exposure Time: More than 120
minutes
Pressurization: 162-200kPa
Depressurization: 13-8kPa
Relative humidity: 40-90%
Aeration is minimum 12 hours
* STERRAD® systems are manufactured by "Johnson & Johnson"
1)
Before performing sterilization, check that the operation data
of sterilizer are in conjunction with min. and max. data
applicable for the probe and the Attachment.
2)
Do not sterilize the probe and the Attachment by Steam
Autoclaving. If you autoclave them, they suffers serious damage
and will be not functional.
The packaging procedure is as follows.
1)
Put the probe into TYVEK pouch.
TYVEK Pouch
Probe
Fig. 14 Packaging in the
pouch
Attachment