Cleaning and Disinfection and Sterilization
The probe and accessory must be reprocessed after each use.
Refer to the reprocessing instruction in this chapter.
WARNINGS
‐
The probe is delivered unsterile. Prior to the
first use, reprocess the probe.
‐
Temperature should not exceed 60°C during
reprocessing.
‐
Probe connector is not water resistant.
Limitations on
reprocessing
The probe is not completely submersible. The
immersible part is shown in Fig.1. The
un-immersible part should be disinfected by wipe
disinfection.
Transportation
before using
The probe should be packed in a sterile pouch
or container to transport from Central Sterile
Supply Department (CSSD) to an operating room.
Be careful not to damage the sterile pouch or
container during transportation.