NOTE
(1) After writing in the USB flash drive is finished, the following screen appears.
If you press the YES button, the run history data stored in the centrifuge is deleted.
If you press the NO button, the run history data stored in the centrifuge is not deleted.
However, if the information about 100 sets of the run parameters of the previous normal
operations has been stored in the centrifuge, the oldest information of the run parameters is
deleted when you perform the centrifuge run.
(2) Be sure to back up your data before using the USB flash drive. Hitachi Koki will not be liable for
any data loss.
(3) This centrifuge cannot use the USB flash drive with security function.
(4) When powered on with the USB flash drive, the centrifuge may not recognize the USB flash
drive. In such a case, pull out the USB flash drive from the centrifuge and then insert it again
into the centrifuge.
Disconnect : LAN communication stops.
LAN communication starts.
Press this button to establish the communication between “himac LogManager” (option)
on your computer and your centrifuge. It is possible to manage operation results of this
centrifuge by using “himac LogManager” (option). If you press this button, either of the
below icons is displayed in the left side of the START button of the Run screen.
Connect :
2-6-4 Administrator (Admin) Functions
[LAN Communication] (option)
If your centrifuge has a LAN communication option,
LAN communication start/stoppage can be specified.
Insert the LAN cable into the LAN port at the right side of the centrifuge.
LAN port
:There is no problem in communication between “himac LogManager” (option) on
your computer and your centrifuge.
NOTE Even if your centrifuge does not have a LAN communication option, the Connect button is
surrounded with a red frame as you press the Connect button. In that case, this icon is
always displayed in the Run screen, but there is no problem.
:There might be problems in communication between “himac LogManager” (option)
on your computer and your centrifuge. Follow the below instructions.
Check whether the LAN cable is connected to the centrifuge properly.
Check whether the “himac LogManager” on your computer is started.
Check whether the LAN works properly.
The following message is displayed.
“Transmission was completed. Do
you want to delete run history data
stored in centrifuge?”
2-68