2-42
Step
Touch panel operation
Screen displays and notes
4
Press the Enter button.
5 Press the START button.
○
The START time and STOP time for centrifugation are
displayed in the RTC area by pressing the START
button. This centrifuge will turn the rotor automatically
at the specified time.
○
Pressing the START button enters the system into
a condition causing the centrifuge to wait until the set
time comes. This centrifuge will turn the rotor
automatically at the specified time and run it for the
specified time.
○
Once the RTC setting is made, you cannot change the
run time (centrifugation time). If you wish to make a
change to the run time, cancel RTC first. If you want to
cancel the RTC, press the RTC button in the Run
screen and press the Cancel button in the screen
for setting RTC.
○
If you cancel RTC when the user lockout function (see
Section 2-6-4 (2) on this function) is enabled, the
following screen appears. Inputting the “login user
PIN” or the “administrator PIN” is required in this
screen.
2-3-5 RTC (real-time control) Operation
The RTC operation does
not get activated unless
you press the START
button.
The START
confirmation box
appears.
If you want to cancel the
RTC operation, press the
STOP button.