9
Installing the Instrument
Adjuster nut
M16 nut
Adjuster
C
D
B
A
1
Move the instrument to the desired
installation location.
2
Loosen the M16 nuts using a 24 mm
wrench.
3
Place a level at the center of the work
table (shaded area in the figure).
4
Turn the adjuster nuts of adjusters A,
C, and D and set the adjusters on the
floor. Turn the adjuster nuts twice (720°)
in the direction in which the adjusters
stick out.
Level along the X-axis with adjusters C and D
and along the Y-axis with adjusters A and D.
5
Adjust so that the level readings are
within 5 seconds of level.
6
Tighten the M16 nuts and lock the
adjusters.