103
Error display
6
6.4 Error display
Window display when an error occurs
Display
Cause and solution
<Emergency Stop Conditions>
•
Emergency button pressed,
malfunction or wire disconnection
•
External interlock
•
Operation Selector switched
The following causes are suspected.
•
Measurement stopped because the emergency stop button was
pressed or an interlock signal was externally input.
→ Press the
[Reset]
button of the Z5017 PLC Rack to cancel the
error.
•
The instrument failed or is disconnected.
→ Contact your authorized Hioki distributor or reseller.
Tape switch was pressed.
The tape switches of the stage were touched. Press the
[Reset]
button to cancel the error.
Tape switch was disconnected.
The tape switches are disconnected.
Contact your authorized Hioki distributor or reseller.
Possibility of the problem of PLC.
There may be problems inside the PLC.
Contact your authorized Hioki distributor or reseller.
Disconnection of a Cable
The following causes are suspected.
•
The LAN cable is not connected.
•
The PLC is not in operation.
Check the LAN cable connection. If there is no problem with the
LAN cable connection, the PLC may have failed.
Contact your authorized Hioki distributor or reseller.
The detection of an axis limit.
The stage is outside of the movable range.
Operate the pendant and return the stage to the movable range.
Reference: “3.19 Positioning a Workpiece” (p. 65)
Low Battery of PLC (or Servopack).
The battery charge is low.
Replace with new batteries (model: JZSP-BA01) within two weeks.
Reference: “Replacing batteries for PLC and servo pack” (p. 104)
Summary of Contents for TS2400
Page 2: ......
Page 16: ...12 Product Overview ...
Page 34: ...30 Model Z5018 Switch Box ...
Page 74: ...70 Connecting the Measuring Instrument and a Workpiece ...
Page 82: ...78 Automatic measurement ...
Page 112: ...Appx 2 ...
Page 113: ......
Page 114: ......
Page 115: ......
Page 116: ......