7.4 MR8904 CAN Editor General Specifications
61
Transmission settings
1. Transmission number: 1 to 16
2. Operating mode: Interval, Response, One Shot, One Shot Program, Repeat
Program
3. CAN output port settings: Port 1, port 2
4. Frame types: Data frame, remote frame
5. Transmission ID: 1 to 1FFFFFFF H
6. Transmission byte length: 0 to 8
7. Transmit data: (Transmit byte length) bytes of data are set using hexadecimal
notation.
8. Response ID: 1 to 1FFFFFFF H (response transmission only)
9. Transmission interval: 1 to 9,999 (× 10 ms) (Interval, One Shot Program, and
Repeat Program only)
Communications with Memory
HiCorder
1. Search for USB-connected Memory HiCorder; when multiple instruments are
connected, select one.
2. Send register list, CAN communications settings, analog channel settings, logic
channel settings, transmission settings, and display upper and lower limit values
to the Memory HiCorder.
3. Receive register list, CAN communications settings, analog channel settings,
logic channel settings, transmission settings, and display upper and lower limit
values from the Memory HiCorder.
Display functions
1. CAN definition data list display: Displays all CAN definition data settings.
2. Register list display: Displays the item number, input port number, signal name,
label, unit, data length, and display upper and lower limit values.
3. Unit position: Displays an asterisk (*) at positions in the connected device where
no MR8904 is present.
4. CAN communications settings display: Displays all CAN communications set-
tings.
5. Analog assign list display: Displays the channel, input port number, signal
name, label, unit, and display upper and lower limit values.
6. Logic assign list display: Displays the channel, input port number, signal name,
label, and bit position.
7. Send list display: Displays all transmission settings.
Print functions
1. Print register list: Prints the item number, input port number, signal name, label,
unit, and data length.
2. Print CAN communications settings: Prints all CAN communications settings.
3. Print analog assign list: Prints the channel, input port number, signal name,
label, and unit.
4. Print logic assign list: Prints the channel, input port number, signal name, label,
and bit position.
5. Print send list: Prints all transmission settings.
Save functions
1. CAN definition data: Binary format (extension of .CDF, compatible with the soft-
ware for Model 8910)
2. Settings data: Binary format (extension of .CES, all settings except CAN defini-
tion data)
Summary of Contents for MR8904
Page 1: ...MR8904 Instruction Manual CAN UNIT EN Sept 2018 Revised edition 3 MR8904A981 03 18 09H...
Page 2: ......
Page 10: ...Usage Notes 6...
Page 14: ...10...
Page 24: ...3 3 Connecting the Unit to a PC 20...
Page 66: ...7 4 MR8904 CAN Editor General Specifications 62...
Page 69: ......
Page 70: ......
Page 71: ......
Page 72: ......