![Hioki MR8904 Instruction Manual Download Page 61](http://html1.mh-extra.com/html/hioki/mr8904/mr8904_instruction-manual_2137100061.webp)
7.1 Model MR8904 Specifications
57
Specifications
Chapter 7
7.1
Model MR8904 Specifications
Compliant CAN standards
ISO 11898 CAN 2.0b
ISO 11898-1 CAN protocol
ISO 11898-2 High-speed physical layer for CAN
ISO 11898-3 Low-speed fault-tolerant physical layer for CAN
SAE J2411 Single wire physical layer for CAN
No. of CAN ports
2 (Port 1, Port 2)
Input CAN connectors
Two D-sub 9-pin male connectors
CAN interfaces
User-selectable high-speed CAN, low-speed CAN, or single-wire CAN for each port
(built-in compatible transceiver) from the MR8904 CAN Editor
ACK On/Off
ACK transmission can be turned on and off for CAN signal reception with MR8904
CAN Editor.
Terminators
Built-in 120 ±10
terminator; on/off setting with commands
Communications speeds
High-speed CAN : 1M, 800k, 500k, 250k, 125k, 100k, 83.3k, 62.5k, 50 kbps
Low-speed CAN : 125k, 100k, 83.3k, 62.5k, 50k, 33.3k, 20k, 10kbps
Single-wire CAN : 83.3k, 62.5k, 50k, 33.3k, 20k, 10kbps
Baud rate settings can be made either with the MR8904 CAN Editor or using auto-
matic detection.
Signal resolution output
channels
15 channels equivalent to 16-bit analog signal
16 1-bit channels equivalent to logic signal
Target signal format
1-bit signals
: Use 1 logic channel or 1 analog channel.
1- to 16-bit signals
: Use 1 analog channel.
17- to 32-bit signals : Use 2 analog channels.
Signals over 32 bits : Not supported.
ID trigger
When the set ID is received, a high-level pulse is output to the specified logic chan-
nel. The pulse width is 50
s if the time axis is within 5 ms/div or 1 sample if the time
axis is more than 10 ms/div.
Response time
Within 200
s of CAN message reception completion
CAN message transmission
The set CAN message is transmitted to the bus for each port.
One Shot : Transmission once only
Interval
: Transmission at the set interval (10 ms to 99.99 s)
Response : Transmission when the set ID is received
Program : Transmission in the set order at the set interval (10 ms to 99.99 s)
Transmission timing depends on the start time and message transmission key.
Status indicator LEDs
Separate for ports 1 and 2
Turn green when a CAN signal with the set ID is received.
Turn red when an error occurs.
Operating temperature and
humidity
As per the Memory HiCorder into which the MR8904 has been installed
Operating environment
As per the Memory HiCorder into which the MR8904 has been installed
Storage temperature and
humidity
Temperature : -20 to 60°C (-4 to 140°F)
Humidity
: -20°C to 40°C (-4 to 104°F) 80% RH or less (no condensation)
40°C to 45°C (104 to 113°F) 60% RH or less (no condensation)
45°C to 60°C (113 to 140°F) 50% RH or less (no condensation)
Dimensions
Approx. 119.5W×18.8H×151.5D mm (4.70”W x 0.74”H x 5.96”D)
Product warranty period 3 years
Summary of Contents for MR8904
Page 1: ...MR8904 Instruction Manual CAN UNIT EN Sept 2018 Revised edition 3 MR8904A981 03 18 09H...
Page 2: ......
Page 10: ...Usage Notes 6...
Page 14: ...10...
Page 24: ...3 3 Connecting the Unit to a PC 20...
Page 66: ...7 4 MR8904 CAN Editor General Specifications 62...
Page 69: ......
Page 70: ......
Page 71: ......
Page 72: ......